Ordering information
ESA612 United States, 115 V 20 A
ESA612-02 Europe, 230 V
ESA612-01 France, 230 V
ESA612-03 Israel, 230 V
ESA612-05 Australia, 230 V
ESA612-06 United Kingdom, 230 V
ESA612-07 Switzerland, 230 V
ESA612-08 Thailand, 230 V
ESA612-09 Japan, 100 V
ESA612-10 North America, 220 V
TA-ESA612-US United States, 115 V
20 A w/Test Automation
TA-ESA612-EUR Europe, 230 V
w/Test Automation
TA-ESA612-FR France, 230 V
w/Test Automation
TA-ESA612-ISR Israel, 230 V
w/Test Automation
TA-ESA612-AUS Australia, 230 V
w/Test Automation
TA-ESA612-UK United Kingdom,
230 V w/Test Automation
TA-ESA612-SWI Switzerland, 230 V
w/Test Automation
TA-ESA612-THAI Thailand, 230 V
w/Test Automation
TA-ESA612-JAPAN Japan, 100 V
w/Test Automation
TA-ESA612-NA220V North America,
220 V, w/ Test Automation
Included accessories
CD-ROM Operator’s Manual
(multilingual CD)
MANUAL Getting-Started Guide (hard
copy, multilingual)
CABLE ASSEMBLY Data Transfer Cable
ESA612 Accessory Kit (country
specific)
2719-0154 15 A to 20 A Adapter
(US only)
esa620-npa Null Post Adapter
esa612-2016 5-to-5 Banana Jack to
ECG (BJ2ECG) Adapter
9530-0075 Carry Case
Detachable Power Cord (country
specific)
Included accessories for ESA612
with test automation
All of the above, plus:
ANSUR ESA612 Ansur Test-
Automation Software ESA612 plug-in
Optional accessories
1903307 Retractable Test Leads
2242165 Ground Pin Adapter (US
receptacle testing ground lug)
3392119 1210 Adapter Box Assembly
3454829 Ansur ESA612 Plug-In
License Key
ESA612
Electrical Safety Analyzer
Specications
Frequency response/
accuracy
DC to 1 kHz ± (1 % of reading + (1 μA or 1
LSD, whichever is greater))
1 kHz to 100 kHz ± (2 % of reading + (1 μA or 1
LSD, whichever is greater))
1 kHz to 5 kHz
(current > 1.6 mA)
± (4 % of reading + (1 µA or 1
LSD, whichever is greater))
100 kHz to 1 MHz ± (5 % of reading + (1 μA or 1
LSD, whichever is greater))
Note: Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are
+ (2.5 µA or 1 LSD, whichever is greater)
Leakage tests Ground wire (earth), Chassis (enclosure), Lead to ground (patient),
Lead to lead (patient auxiliary), Lead isolation (mains on applied
part), Direct equipment , Direct applied part, Alternative equip-
ment, Alternative applied part, Point to point
Mains on applied part
test voltage
100 % of mains
Differential leakage
Ranges 75 µA to 199 µA
200 µA to 2000 µA
2 mA to 20 mA
Accuracy ± 10 % of reading + (2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges
0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ
Accuracy
± (2 % of reading + 0.2 MΩ)
± (7.5 % of reading + 0.2 MΩ)
Source test voltage 500 V dc
250 V dc
Insulation resistance tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed
accessible conductive part) and AP- NE (non-earthed accessible
conductive part)
ECG performance waveforms
Accuracy ± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration
Waveforms: rates
Ventricular fibrillation
ECG complex (BPM): 30, 60, 120, 180, and 240
Square wave (50 % duty cycle) (Hz): 0.125 and 2
Sine wave (Hz): 10, 40, 50, 60, and 100
Triangle wave (Hz): 2
Pulse (63 ms pulse width): 30 BPM and 60 BPM
Power ratings
Mains voltage outlet 120 V ac or 230 V ac
Mains voltage inlet
power range
90 to 132 V ac rms 180 to 264 V ac rms
Maximum current 20 A 16 A
Hz 50 or 60 50 or 60
Physical case
Dimensions (L x W x H) 17.63 cm x 8.38 cm x 28.45 cm (6.94 in x 3.30 in x 11.20 in)
Weight 1.6 kg (3.5 lb)
Environmental specifications
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 90 % non-condensing
Altitude 120 V ac mains supply voltage up to 5,000 m,
230 V ac mains supply voltage up to 2,000 m
General
Warranty Two-year extended warranty (no-cost, available after first-year
calibration at the Fluke Biomedical Cleveland Service Center,
otherwise standard one-year warranty applies)
Electrical Safety Analyzers