40 Multi-Function Simulators
NIBP simulations Pulse 2 mmHg max into 500 ml NIBP system
Volume of air moved 1 ml max
Simulations Adult: 60/30 (40), 120/80 (93); 150/100 (117);
and 200/150 (167)
Neonatal: 35/15 (22) and 70/40 (50)
Repeatability Within ± 2 mmHg (at maximal pulse size inde-
pendent of device under test)
Synchronization To ECG heartrate (maximal rate 120 BPM)
Leak test Target pressure 20 mmHg to 400 mmHg
Elapse time 0:30 minutes to 5:00 minutes: seconds in 30
second steps
Leakage rate 1 mmHg/minute to 200 mmHg/minute
Pressure relief test range 100 mmHg to 400 mmHg
Presets and autosequences
Presets Normal
Hypertensive
Hypotensive
Autosequences Cardiac failure sequence
Exercise sequence
Respiration sequence
Monitor testing sequence
ProSim 4
Vital Signs Simulator
General specifications
Temperature Operating 10 °C to 40 °C (50 °F to 104 °F)
Storage -20 °C to +60 °C (-4 °F to + 140 °F)
Humidity 10 % to 90 % non-condensing
Altitude 3,000 meters (9,843 ft)
Dimensions (L x W x H) 18 cm x 9.3 cm x 5.5 cm (7.1 in x 3.7 in x 2.2 in)
Display LCD touch-screen color display
Communication USB port (for calibration and firmware updates only)
Power Lithium-ion rechargeable battery
Battery charger TBD
Battery life Four hours (minimum), 40 NIBP cycles typical
Weight 0.88 kg (1.93 lb)
Safety standards IEC 61010 -1:2001
Certifications CE, CSA, C-TICK N10140, RoHs
Electromagnetic compatibility (EMC) I EC 61326-1:2006