62
Optional accessories
5000C-02 Performance Enhancement
Module 2: RS-232/printer
5000C-03 Performance Enhancement
Module 3: 100 records
5000C-04 Performance Enhancement
Module 4: Expanded memory
5000C-05 Performance Enhancement
Module 5: Waves/extended test
5000C-06 Performance Enhancement
Module 6: Data transfer
5000C-07 Performance Enhancement
Module 7: medCheck
5000C-08 Performance Enhancement
Module 8: Defibrillator
5000C-09 Performance Enhancement
Module 9: IV pump
5000C-10 Performance Enhancement
Module 10: CMMS interface
5000C-11 Performance Enhancement
Module 11: ESU
5000C-12 Performance Enhancement
Module 12: Sp0
2
5000C-13 Performance Enhancement
Module 13: Pacer
5000C-14 Performance Enhancement
Module 14: NIBP
5000C-WEDGE Wedge Adapter (eight
25 in serial ports, as well as AT
or PS/2 keyboard port)
9513-0212 Mini PC-style External
Keyboard (83 keys, AT or PS/2, wedge
adapter required
The medTester 5000C is an
automated system designed for
electrical safety testing and
performance verification. It is
compatible with most Fluke
Biomedical testing devices
and a majority of the popular
Computerized Maintenance
Management Systems (CMMS)
in the US. The medTester
5000C provides a completely
integrated solution for stan-
dardized and streamlined
testing and record keeping.
Modes of operation Fully equipped, with four operational modes: manual,
autosequence, medCheck, and remote control
Input power supply Line voltage/frequency input: 115 V ac ± 10 %/60 Hz
Test-receptacle type USA, 20 A
System/line voltage
Range (full scale) 200 V
Accuracy ± 5 % of range ± 1 LSD
Resolution 0.1 V
Equipment current
Range (full scale) 0 A to 20 A
Accuracy ± 5 % of range
Resolution 0.01 A
Ground resistance
Range (full scale)
0 Ω to 2 Ω
Accuracy ± 1 % of range
Resolution
0.001 Ω (1 mΩ)
Current source 100 mA dc
Measurement type True four-terminal technique
Test leads Kelvin (2) insulated clip
Leakage-current/voltage gradient
Ranges (full scale) 200 µA and 2000 µA or mV
Accuracy ± 1 % of reading DC + 3 LSD from 48 Hz to 1 kHz
± 2.5 % of reading + 3 LSD from 1 KHz to 100 KHz
± 5 % + 3 LSD from 100 KHz to 1Hz
Resolution 0.1 µA or 0.1 mV
Measurement type True-rms (autoranging) (ac + dc or dc only response)
Test-load selection ANSI/AAMI ES1 1993
Test-load impedence
1000 Ω ± 1 % at dc
Isolation test
Test selection (full scale) Patient leads to ground
Lead combinations All leads; or individual leads RL, RA, LA, LL, and V1/V6
(V1 through V6 tested as a single lead)
Available current
Limited by internal 120 kΩ resistor
Resolution 0.1 µA
Ranges (full scale) 200 µA and 2000 µA
medTester 5000C
Automated Biomedical Equipment Test System
Specications
Key features
• Easy verification of biomedical
equipment to manufacturer’s
specifications
• Ten preprogrammed and five
user-programmable electrical-
safety-testing sequences
• Convenient transfer of
equipment inventory and
testing procedures from CMMS
• Module options to automate
testing of most Fluke
Biomedical testing devices
• Automatic storage of detailed
test results for printing or
transfer to CMMS
• Compliant with ANSI/AAMI
(1993) and NFPA-99 (2005)
standards
• 20 A device testing with GFCI
protection
• Wedge hardware option for
extended serial port use,
optional PC-style keyboard and
barcode scan gun
Automation Solutions